This product is for Professional Use only and not for sale to the general public for home use.
Please log in with a business account to purchase.
PYROSTAR ES-F Multi Test Endotoxin Testing Kit without Control Standard Endotoxin (CSE) GCA method and 0.015EU/mL KTA Sensitivity 1 pack of 100x5.2mL
Product code: WPEM-50015
Pack Size: 100x5.2mL
CE Verified: Research Use Only
Contact us
This is not a regular stock item and therefore not subject to our standard delivery schedule. Please contact us on +44 (0) 2380 483000 for advice on expected delivery date.
Product Information
PYROSTAR is Fujifilm Wako’s range of endotoxin testing instruments and reagents, based on the Limulus Amoebocyte Lysate (LAL) assay.
The assay operates using lysate isolated from the blood of the Atlantic Horseshoe Crab (Limulus polyphemus). When exposed to bacterial endotoxins, generally arising from lipopolysaccharides (LPS) in the bacterial cell wall, these lysates coagulate and in turn cause the crab’s blood to clot. Pyrostar's LAL assay harnesses this to provide the most sensitive endotoxin detection method approved by the FDA.
This PYROSTAR ES-F Multi Test Endotoxin Testing Kit provides 4 vials of PYROSTAR ES-F lyophilised reagent. Each vial allows for multiple tests using either Gel Clot Assay (GCA) or Kinetic Turbidimetric Assay (KTA) methods, with a sensitivity of 0.015EU/mL. To avoid cross-reactivity with other cell wall components, such as β-1,3-glucan, ES-F also contains carboxymethylated curdlan.
Gel Clot Assay (GCA) Method
The Gel Clot Assay is the simplest endotoxin detection method and being qualitative, does not require a photometric instrument, such as the Fujifilm Wako ET-7000 Toxinometer. This makes the assay much cheaper to run as the toxinometer can be replaced by a 37°C water bath.
The assay works by incubating samples with the ES-F reagent in endotoxin-free tubes for 60 minutes at 37°C and inverting the tubes to test for coagulation. If coagulation has not occurred, the sample can be confirmed to be endotoxin free up to 0.015EU/mL. However, if coagulation occurs, endotoxins are present and by further testing using serial sample dilutions, the endotoxin level can be mathematically approximated.
Kinetic Turbidimetric Assay (KTA) Method
The KTA method is a quantitative method for detection of bacterial endotoxins, using a photometric instrument, such as the ET-7000 Toxinometer. It utilises the same coagulation principle as the GCA, however by using a toxinometer to continually measure sample turbidity, coagulation speed can be determined and through regression analysis in comparison to a standard curve, the toxinometer will calculate the sample endotoxin level.
Pack Contents:
100 vials ES-F Reagent
• Reconstitution: Ensure lysate is at bottom of vial by knocking on a hard surface and add 5.2mL of LAL Reagent Water (LRW) (Not included)
• Unreconstituted Storage: 2-10°C
• Reconstituted Storage: 2-8°C (≤4 hours), -15°C (≤14 days)
• Storage (Reconstituted): 2-10°C (≤1 Month). Do not store below freezing
Not Included (Product Code(s) where available):
•Reference Standard Endotoxin (RSE): USP Endotoxin Reference Standard
•Control Standard Endotoxin (CSE) (CSE4037-5006)
•LAL Reagent Water (LRW) (12x100mL LRW-12100, 20x30mL LRW-2030, 12x125mL LRW-12125)
•Endotoxin free pipette tips (200µL 291-35021, 1mL 298-35031, 200µL Ext 294-35011)
•Endotoxin Free Dilution Tubes (200 Small CT-1075, 1250 Small CT-1075B, 50 Large DL-13100)
•ET-7000 Toxinometer (Set 299-35821, Single Module 292-35811)
•Water Bath
•Endotoxin Free Test Tubes (293-26551) and aluminium caps(293-28251). (Both:292-32751)
The assay operates using lysate isolated from the blood of the Atlantic Horseshoe Crab (Limulus polyphemus). When exposed to bacterial endotoxins, generally arising from lipopolysaccharides (LPS) in the bacterial cell wall, these lysates coagulate and in turn cause the crab’s blood to clot. Pyrostar's LAL assay harnesses this to provide the most sensitive endotoxin detection method approved by the FDA.
This PYROSTAR ES-F Multi Test Endotoxin Testing Kit provides 4 vials of PYROSTAR ES-F lyophilised reagent. Each vial allows for multiple tests using either Gel Clot Assay (GCA) or Kinetic Turbidimetric Assay (KTA) methods, with a sensitivity of 0.015EU/mL. To avoid cross-reactivity with other cell wall components, such as β-1,3-glucan, ES-F also contains carboxymethylated curdlan.
Gel Clot Assay (GCA) Method
The Gel Clot Assay is the simplest endotoxin detection method and being qualitative, does not require a photometric instrument, such as the Fujifilm Wako ET-7000 Toxinometer. This makes the assay much cheaper to run as the toxinometer can be replaced by a 37°C water bath.
The assay works by incubating samples with the ES-F reagent in endotoxin-free tubes for 60 minutes at 37°C and inverting the tubes to test for coagulation. If coagulation has not occurred, the sample can be confirmed to be endotoxin free up to 0.015EU/mL. However, if coagulation occurs, endotoxins are present and by further testing using serial sample dilutions, the endotoxin level can be mathematically approximated.
Kinetic Turbidimetric Assay (KTA) Method
The KTA method is a quantitative method for detection of bacterial endotoxins, using a photometric instrument, such as the ET-7000 Toxinometer. It utilises the same coagulation principle as the GCA, however by using a toxinometer to continually measure sample turbidity, coagulation speed can be determined and through regression analysis in comparison to a standard curve, the toxinometer will calculate the sample endotoxin level.
Pack Contents:
100 vials ES-F Reagent
• Reconstitution: Ensure lysate is at bottom of vial by knocking on a hard surface and add 5.2mL of LAL Reagent Water (LRW) (Not included)
• Unreconstituted Storage: 2-10°C
• Reconstituted Storage: 2-8°C (≤4 hours), -15°C (≤14 days)
• Storage (Reconstituted): 2-10°C (≤1 Month). Do not store below freezing
Not Included (Product Code(s) where available):
•Reference Standard Endotoxin (RSE): USP Endotoxin Reference Standard
•Control Standard Endotoxin (CSE) (CSE4037-5006)
•LAL Reagent Water (LRW) (12x100mL LRW-12100, 20x30mL LRW-2030, 12x125mL LRW-12125)
•Endotoxin free pipette tips (200µL 291-35021, 1mL 298-35031, 200µL Ext 294-35011)
•Endotoxin Free Dilution Tubes (200 Small CT-1075, 1250 Small CT-1075B, 50 Large DL-13100)
•ET-7000 Toxinometer (Set 299-35821, Single Module 292-35811)
•Water Bath
•Endotoxin Free Test Tubes (293-26551) and aluminium caps(293-28251). (Both:292-32751)
Specification
- Brand FUJIFILM Wako Chemicals Europe GmbH – Lab Chem
- Sterile Y
- CE Certified Research Use Only
- Warranty Period Standard Terms Apply
- Is product classified as dangerous goods? No
- Does the product contain latex? Unspecified
- Detection Type Limulus Amebocyte Lysate (LAL) Assay
- Contents 100 x 5.2mL PYROSTAR ES-F Multi-Test Vials (Sensitivity 0.015EU/mL)
- Specificity Bacterial Endotoxin
- Time Required 1 Hour
- Application Bacterial Endotoxin Testing
Storage Details
- Pack Description 100x5.2mL
- Shipping Condition Blue Ice
- Storage Condition Fridge
- Pack Length (m) 0.5
- Pack Width (m) 0.5
- Pack Height (m) 0.1
- Pack Weight (kg) 0.00
You may also be interested in