Bioer GeneTouch™ Thermal Cycler

GeneTouch™ thermal cyclers are flexible and robust, with interchangeable blocks to support busy labs and changing workflows.

With 5 interchangeable blocks, 2 modes of temperature control, adjustable temperature ramping, incremental/decremental time setting, adjustable height and temperature hot-lid and the ability to store programmes on the instrument, on USB memory flash drives or on your PC, this cycler is suitable for a multitude of PCR applications and incubations.

Key Features

Versatile PCR Setups

Rapid block changing mechanism accommodating high throughput.

Great Performance

Gradient setting tool for rapid optimised configuration, along with tube or block temperature control modes.

Optimised PCR

Adjustable hot-lid temperatures and height enabling oil-free operation with pressure adjustable fit, and automatic shut-off.

Flexible Integration

Multiple units able to be programmed and ran from a LAN network, direct from a PC, or as stand-alone instruments.

Multi-user Functionality

Inbuilt program storage of up to 250, which can be expanded indefinitely via a USB drive.

GeneTouch Specifications

Specifications
Block TC1110 TC1120 TC1130 TC1140 TC1150
Block Sample Capacity 96-well 48-well 48-well universal block 384-well 4 x standard in situ plates
Cooling Technology Peltier elements
Display 6.5 inch, 262k-colour LCD display and touch screen
Temperature Range 4 - 105ºC
Heating speed of ramping Up to 4ºC/sec Up to 4ºC/sec Up to 2.8ºC/sec Up to 2.8ºC/sec Up to 1.8ºC/sec
Cooling speed of ramping Up to 4ºC/sec Up to 4ºC/sec Up to 2.8ºC/sec Up to 2.8ºC/sec Up to 1.8ºC/sec
Temperature uniformity ≤±0.2ºC
Temperature accuracy ≤±0.1ºC (55ºC) : ≤±0.2 C (≥90ºC)
Temperature control modes Block or Tube
Memory

250 programs

Limitless with USB flash drive storage
Maximum number of cycles 99
Gradient range of temp differential 1 - 30ºC
Hot-lid temperature range 30 - 110ºC
Interfaces USB and LAN
Power supply 100-240v, 50-60Hz, 600W
Dimensions (WxDxH) 368 x 250 x 285mm
Net weight 10.5kg

NOTE: Stated values were obtained under controlled laboratory conditions